<% ProductGroup=0 Product="" Industry=0 Author="" Title="" LastUpdate=now() Country="WW" PageType=21 NavigationOverride=1 BodyGraphic="" ExtraNav="" %>

Software Release Bulletin


OpenLAB CDS ChemStation Edition     M83xxAA

Apr 5 2011 8:03AM

Preface
-------------------------
This Software Release Bulletin documents all fixes and enhancements that are incorporated the new release.
-------------------------


Keyword Glossary


Reports

Keyword Index


Keyword: Reports

Extended performance report (IR) reports "Infinity" for plates per meter
KPR#:1581
System may hang when ".DIF" output is choosen as report type.
KPR#:1585
n/a
KPR#:1610


KPR#: 1581    Product: Chemstation     ChemStation (G2070BA)    n/a

Keyword: Reports

One-line Description:

Extended performance report (IR) reports "Infinity" for plates per meter

Problem:
In case when no column information is available report shows"infinity" instead of the expected "--".
Temporary Solution:
n/a
Fix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02

KPR#: 1585    Product: Chemstation     ChemStation (G2070BA)    n/a

Keyword: Reports

One-line Description:

System may hang when ".DIF" output is choosen as report type.

Problem:
Due to a memory leak in the DIF file creation component, when long sequences are running and the ".DIF" output is choosen as report type, the System may hang and abort running instruments. When this occur, the computer needs to be rebooted.
Temporary Solution:
This sporadic hang can be prevented by not selecting ".DIF" output.
Fix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02

KPR#: 1610    Product: Chemstation     ChemStation (G2070BA)    n/a

Keyword: Reports

One-line Description:

n/a

Problem:
'SequenceSummary_Standard.rdl" & ''SequenceSummary_Extended.rdl" reports  do not properly display 'Sequence line' number.
Temporary Solution:
n/a
Fix Information:
This issue is resolved in the software revision C.01.02